Laboratory inverted microscope for research applications. Dye-cast frame, with high stability and ergonomy, for transmitted light and and reflected fluorescence observation.
TECHNICAL SPECIFICATIONS:
Illumination
Transmitted Light:
Light source type X-LED8 with white 8W LED; light intensity control using a knob on left side of the frame.
Color temperature: 6300K
LED average life time approx. 50.000h
Voltage: 110/240Vac, 50/60Hz, 1A ; Fuse: T500mA 250V
Max power required: 13W
Reflected Light:
High efficiency LED, 18W.
Light intensity control using a knob on left side of the frame.
LED average life time approx. 50.000h
Voltage: 110/240Vac, 50/60Hz, 1A ; Fuse: T500mA 250V
Max power required: 24W
Observation Modes
Brightfield, phase contrast, fluorescence B and G.
Focusing
Macro- and micrometric adjustment, with coaxial knobs on both sides of the stand. Adjustable tension.
Stage
Fixed stage, dimensions 250×160 mm.
Glass stage insert with hole for small dimension specimens.
OPTIONAL:
– Mechanical stage mountable on the right side of the stage, total dimension=250×230 mm, X-Y translation range 120×80 mm.
– Metallic interchangeable inserts for slides, Petri dishes, Terasaki, multi-Well plates, etc.
– Pair of side extensions to expand the surface of the stage.
Nosepiece
5 positions, with bidirectional rotation on ball bearings and click stop.
Head
Trinocular observation head, inclined 45°.
Diopter adjustment on left eyepiece.
Interpupillary adjustment 50-75 mm.
Eyepieces
Wide field eyepieces EWF10X/22 with field number 22.
Objectives
Infinity corrected optical system IOS (Infinity Optical System).
Plan-achromatic LWD objectives infinity corrected, for thickness 1.2 mm, made by following objectives:
-) Plan-achromatic IOS LWD 10XPh, N.A. 0.25, W.D. 7.94 mm
-) Plan-achromatic IOS LWD 20XPh, N.A. 0.40, W.D. 7.66 mm
-) Plan-achromatic IOS LWD 40XPh, N.A. 0.60, W.D. 3.71 mm
All objectives are treated with an anti-fungus treatment.
Condenser
LWD condenser, N.A. 0.30, working distance 72 mm.
The condenser can be removed to extend the working distance up to 150 mm.
Epi-fluorescence filterset
Blue and Green fluorescence filtersets:
Fluorescence B: EX 450-490, DM 495, EM 520LP.
Fluorescence G: EX 540-580, DM 585, EM 590LP.
Brightfield filterset
Interferential IF550 green filter.
Dimensions
HEIGHT: 495 mm
WIDTH: 230 mm
WIDTH WITH OPTIONAL MECHANICAL STAGE: 300 mm
DEPTH: 540 mm
WEIGHT: 9 kg
Accessories
Anti-glow fluorescence cap.
Instruction manual and dust cover included.
Transmitted Light:
Light source type X-LED8 with white 8W LED; light intensity control using a knob on left side of the frame.
Color temperature: 6300K
LED average life time approx. 50.000h
Voltage: 110/240Vac, 50/60Hz, 1A ; Fuse: T500mA 250V
Max power required: 13W
Reflected Light:
High efficiency LED, 18W.
Light intensity control using a knob on left side of the frame.
LED average life time approx. 50.000h
Voltage: 110/240Vac, 50/60Hz, 1A ; Fuse: T500mA 250V
Max power required: 24W
Observation Modes
Brightfield, phase contrast, fluorescence B and G.
Focusing
Macro- and micrometric adjustment, with coaxial knobs on both sides of the stand. Adjustable tension.
Stage
Fixed stage, dimensions 250×160 mm.
Glass stage insert with hole for small dimension specimens.
OPTIONAL:
– Mechanical stage mountable on the right side of the stage, total dimension=250×230 mm, X-Y translation range 120×80 mm.
– Metallic interchangeable inserts for slides, Petri dishes, Terasaki, multi-Well plates, etc.
– Pair of side extensions to expand the surface of the stage.
Nosepiece
5 positions, with bidirectional rotation on ball bearings and click stop.
Head
Trinocular observation head, inclined 45°.
Diopter adjustment on left eyepiece.
Interpupillary adjustment 50-75 mm.
Eyepieces
Wide field eyepieces EWF10X/22 with field number 22.
Objectives
Infinity corrected optical system IOS (Infinity Optical System).
Plan-achromatic LWD objectives infinity corrected, for thickness 1.2 mm, made by following objectives:
-) Plan-achromatic IOS LWD 10XPh, N.A. 0.25, W.D. 7.94 mm
-) Plan-achromatic IOS LWD 20XPh, N.A. 0.40, W.D. 7.66 mm
-) Plan-achromatic IOS LWD 40XPh, N.A. 0.60, W.D. 3.71 mm
All objectives are treated with an anti-fungus treatment.
Condenser
LWD condenser, N.A. 0.30, working distance 72 mm.
The condenser can be removed to extend the working distance up to 150 mm.
Epi-fluorescence filterset
Blue and Green fluorescence filtersets:
Fluorescence B: EX 450-490, DM 495, EM 520LP.
Fluorescence G: EX 540-580, DM 585, EM 590LP.
Brightfield filterset
Interferential IF550 green filter.
Dimensions
HEIGHT: 495 mm
WIDTH: 230 mm
WIDTH WITH OPTIONAL MECHANICAL STAGE: 300 mm
DEPTH: 540 mm
WEIGHT: 9 kg
Accessories
Anti-glow fluorescence cap.
Instruction manual and dust cover included.
There are no reviews yet.